July 27, 2020

LS126C UV light intensity meter parameters

LS126C UV light intensity meter parameters
1. Spectral response: λ:(230nm-280nm);λp = 254nm
2. Measuring range: 0-8000 uW/cm2
3. Resolution: 0.1 μW/cm2
4. Measurement accuracy: ±10%
5. Host size: length 130mm × width 70mm × height 28mm
6. Probe size: diameter 45mm×thickness 16mm
7. Probe cable length: 1 meter
8. Instrument weight: about 250 grams
9. Battery: 4 AAA alkaline dry batteries
standard
1. WS/T 367-2012 "Technical Specification for Disinfection in *** Institutions" P15, Appendix A.3 Effect monitoring of ultraviolet disinfection.
2. The 2009 version of "*** Disinfection Technical Specification" P9, 2.3 UV disinfection.
3. Ultraviolet germicidal lamp GB19258-2003 P4, rated value of ultraviolet radiation illuminance.
The above standards stipulate the following:
1. Only 5 minutes after turning on the UV lamp can the UV intensity of the lamp be measured stably.
2. The instrument is placed 1m vertically under the UV tube for measurement.
3. The new ordinary 30W straight tube ultraviolet lamp is qualified if the irradiation intensity is greater than 100uW/cm2.
4. The irradiation intensity of the UV lamp in use must be greater than 70uW/cm2 to qualify.
5. The peak wavelength of the ultraviolet germicidal lamp is 253.7nm. Only the ultraviolet lamp of this wavelength (UVC) has the sterilization function.
6. Different power UV lamps have different radiation ratings.

UV ruler hook
In order to meet the *** standard, the measurement ensures that the UVC ultraviolet irradiance meter is located 1 meter vertically down the middle of the tube. Our company also produces a UV ruler hook for purchase, this hook is 1m long. Place the instrument and hang it directly on the lamp tube to facilitate on-site UV intensity measurement.